Part 2 Isolating and Solving Instrument Problems
"Local Recorder
door open" sta-
tus message
"Local Recorder
out of paper"
status message
"No Local
Recording
(Printhead Tem-
perature)" sta-
tus message
"Local Recorder
disconnected"
status message
"Local Recorder
connected"
"No local alarm
recording avail-
able" status
message
196
Recorder door is open.
The Recorder is availa-
ble, but is not ready for
recording.
The Recorder is out of
paper.
The Recorder is availa-
ble, but is not ready for
recording.
The printhead is too warm
(> 68°C).
The printhead is too
cold(< 0°C).
(For instance, the
Recorder may have been
tranferred from storage,
where the temperature
was below 0°C).
The Recorder is discon-
nected from the Monitor
or switched off
The Recorder is connected
and available.
Local alarm recording has
been selected, but the
Recorder is not available.
Close the door.
Load paper into the
Recorder (refer to the
Operator's Manual for the
XE-50p Chart Recorder
for details.
Switch off the Recorder
and allow it to cool down
for a few minutes, then try
again. If the problem per-
sists, replace the Recorder.
The ambient temperature
in the room where the
Recorder is used should
be above 0°C. Allow the
Recorder to warm up for a
few minutes, then try
again.
If the problem persists,
replace the Recorder.
• Ensure the Recorder
cable is securely
connected to the sockets
on the Monitor and on the
Recorder.
• Ensure that the Recorder
is switched on.
Ensure that the Recorder
is switched on and con-
nected to the Monitor.
Troubleshooting the Instrument