Block Diagram Of The Ecg/Resp; Theory Of Operation - Philips M3001A Service Manual

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Safety
To ensure the safety of the patient, the patient-applied parts are isolated from ground by
optical isolators and a transformer. The circuit is also encapsulated in plastic.

Block Diagram of the ECG/Resp

ECG
Electrodes
From
Patient

Theory of Operation

As ECG and Resp signals pass from the patient to the Monitor, they progress through stages
corresponding to the logical sections of the circuit, as shown in the block diagram. Circuit-
related faults can generally be isolated to one of the stages.
Transducer
Signals are received through patient electrodes and lead cables via the input connector.
Input Protection Network
The Input Protection Network and ESU filter eliminate extraneous signals. This protects the
rest of the circuitry from defibrillator voltages, high frequency interference signals, and
electrostatic discharges.
ECG ASIC
The signals are processed by the ECG Application-Specific Integrated Circuit (ECG ASIC)
which has an input amplifier with a fixed gain for each of the four electrodes. They are then
passed to a digital-to-analog converter (D/A Converter) for offset compensation and then to
an analog-to-digital converter (A/D Converter). The input/output logic (which is controlled
Introduction to the Instrument
C
Input
RA
Protection
LA
Network
LL
RL
Bridge &
Amplifier
Excitation
Current
Source
Electrocardiogram/Respiration (ECG/Resp) Measurement
ECG
ASIC
Demodulator
Respiration
CPU
ROM/RAM
To/From
System CPU
25