Serial Numbers

Serial Numbers
When recording test results, these are always associated with a particular instrument by
means of the serial number. The serial numbers for the Monitor and the Measurement Server
can be seen in the "Revisions" window (press Setup key then select "Revisions"). However,
if a Measurement Server Extension (M3015A or M3016A) is in use, the number will not
appear and must be noted down from the back of the Extension. To do this, remove the
Measurement Server and Extension from the Monitor, so that the back of the extension is
visible.

Passwords

The following table show the passwords you need to enter to access the different modes. Go
to Operating Modes and use the TouchStrip to specify the password. Press OK:

Visual Test

Inspect the system for obvious signs of damage. Also check external leads and accessories.
What to record on the service record:
V:P or V: F
where P=pass, i.e. the system has no obvious signs of damage
and

Power On Test

Step 1
Step 2
What to record on the service record:
142
Demo
Config
42351
13251
F=fail
Switch on the monitor and connect the Measurement Server to the M3046A
Observe whether the monitor boots up successfully without displaying an error code
and if an ECG wave appears on the screen.
Service
25531
Testing the Instrument