Disassembly Procedures for the M8043A Battery Charger and Conditioner

Removing the Cooling Fans

Step 1
Step 2
Step 3
Step 4
240
Open the housing (see "Opening the Housing" on page 239).
Carefully manouvre the baseplate to gain internal access to the left-hand cooling
fan.
Left-hand
Cooling Fan
Remove the four screws holding the cooling fan and grill.
Disconnect the two-pin plug (A) connecting the cooling fan cable to the printed
circuit board. Free the cable from the cable guide (B).