Refit Procedures For The M3015A Measurement Server Extension; Refitting The Co 2 Scrubber - Philips M3001A Service Manual

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Refit Procedures for the M3015A Measurement Server
Extension
Tools Required:
• A thin-bladed screwdriver.
• A pair of large tweezers.
• In addition, for refitting the Pump, you will need a large-bladed screwdriver.
WARNING
There is high voltage inside the Instrument (800V). Do not connect the Measurement
Server Extension to a Monitor while the Extension housing is open.
As well, parts inside the Instrument may be contaminated with bacteria. Protect
yourself from possible infection by wearing examination gloves during these
procedures.
Refitting the CO
WARNING
The CO
not open or damage the CO
material, flush the area immediately with water and consult a doctor.
To refit the CO
Step 1
Step 2
Step 3
Repairing the Instrument

Refit Procedures for the M3015A Measurement Server Extension

Scrubber
2
Scrubber contains lithium hydroxide monohydrate. This is a strong base. Do
2
Scrubber. If you come into contact with the CO
2
Scrubber, do the following:
2
Feed the end of the CO
Scrubber through the bracket to meet the Extension intake
2
tube.
Push the intake tube firmly into the scrubber end to connect it.
Holding the body of the CO
air intake under the second bracket and position it.
Scrubber with tweezers, feed the CO
2
Scrubber
2
Scrubber fresh
2
237