System Self-Test Values - Philips M3001A Service Manual

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

System Self-Test Values

Module
ECG and ECG/
Resp
Pressure
SpO
Temperature
Pleth Pulse Rate
NBP
Sidestream CO
Mainstream CO
Testing the Instrument
Test Numeric Limits
100 bpm in ADULT mode
ECG
125 bpm in NEO/PEDI mode
15 rpm in ADULT mode
Resp
30 rpm in PEDI mode
55 rpm in NEO mode
a
Systolic
120 mmHg ADULT
60 mmHg PEDI/NEO
Diastolic
0 mmHg ADULT/PEDI/
NEO
a
/Pleth
777
2
9
SpO
100%
2
Pleth
Wave on display
a
40 °C (104 °F)
60 bpm in ADULT/NEO
mode
120/80 (90) ADULT
100/60 (80) PEDI
80/50 (60) NEO
b
EtCO
40 mmHg
2
2
ImCO
2
AWRR 20 rpm
a
EtCO
40 mmHg
2
2
ImCO
2
AWRR 20 rpm
a. An appropriate transducer must be plugged into the Server or Server Extension in
order to get the test reading.
b. Appropriate tubing must be plugged into the Server Extension in order to get the
test reading. When in Neonatal mode the test signals are switched on, a "CO
Equip. Malf." message will appear until the test signals are switched off; no
action is required.
Test Waveform
Simulated squarewave and
numeric
Simulated Resp wave and numeric
Simulated squarewave and
numerics
Numeric only
Simulated wave
Numeric only
Simulated squarewave and
numerics
Simulated squarewave and
0 mmHg
numerics
Simulated squarewave and
0 mmHg
numerics
Functionality Assurance Tests