M3/m4 monitors,measurement server,measurement server extensions.
Hide thumbs

Advertisement

Message
SpO
TEMP EQUIP MALF
CO
BAD SERVER LINK
BAD SERVER LINK
plus
"Measurement Server
Revision not supported"
status message in red.
"Some measurements are
not supported by the
Monitor" prompt message
Troubleshooting the Instrument
EQUIP MALF
2
EQUIP MALF
2
Troubleshooting the Installed Instrument
What To Do
Measurement defective. Exchange the
Measurement Server.
The Measurement Server should be investigated
technically as soon as possible.
Measurement defective. Exchange the
Measurement Server or Measurement Server
Extension.
The Measurement Server or Measurement Server
Extension should be investigated technically as
soon as possible.
Measurement defective. Exchange the
Measurement Server Extension or (for mainstream
CO
) the transducer.
2
You cannot use this combination of Monitor,
Measurement Server and cable.
Switch off the monitor and contact your biomedical
department.
An M3000A Measurement Server with revision A
software is connected to an M3046A Monitor with
a software revision of B or later. This combination
does not allow monitoring.
A measurement extension (M3015A or M3016A) is
connected to an M3046A M3 monitor.
No measurements (for example CO
Temp) from the Measurement Extension are
available.
If the extra measurements are required, consider
using a monitor that supports these measurements.
, 2nd Press/
2
211