Part 2 Isolating and Solving Instrument Problems
"No Central -
duplicate moni-
tor label"
prompt message
"Assigned Cen-
tral is not avail-
able" prompt
message
"No Central -
software incom-
patible" prompt
message
"Central can not
identify this
bed" prompt
message
Measurement Server related problems
No waves.
No ECG data or
waveform dis-
played when
both defib sync
and balloon
pump are done
together.
190
Monitor label configura-
tion is in conflict with
another monitor
The Information Center is
switched off or not acces-
sible
The location of this bed is
not clear.
Measurement Server
defective.
LEDs on display unit
blink but Server LED does
not.
2-circuit cable used in a 3-
circuit ECG output jack.
Check that the monitor
label assigned to this mon-
itor has not been changed
locally. Check the assign-
ment at the Information
Center
Ask Biomed department
to check software revi-
sions.
Check the bed assign-
ments at the Information
Center.
Check green LED on
Server. Exchange the
Server.
Connection between the
Server and the display is
lost. Exchange the Server.
Use a bud box when set-
ting up ECG.