Part 1 Troubleshooting Checklists
Data Flow Marker In and ECG Wave
The following illustrates the data flow for Marker In and ECG Wave:
Defib <—> Monitor <—> Measurement Server:
176
Defib
Cable
ECG Wave
Marker
DEFIB
Marker In
Data Path
ECG Wave
Data Path
ECG_OUT
Isolat.
CPU,Communication
System and Video
Link Bar
Connector
RS485
LAN
Link Bar
Server
Application
Software
Troubleshooting the Instrument
Display
Assembly
Link Bar
Marker merged to
ECG Wave here