Philips M3001A Service Manual page 203

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

"Measurement
Server Configu-
ration not sup-
ported" status
message
"Some measure-
ments in MMS
Extension are
not supported"
prompt message
Troubleshooting the Instrument
Part 2 Isolating and Solving Instrument Problems
Either
A Measurement Extension
(M3015A or M3016A) is
connected to a standard
M3000A Measurement
Server (noninvasive meas-
urements only) and an
M3046A Monitor with
software Rev.D or lower.
No measurements (CO
2nd Press/Temp) from the
Measurement Extension
are available.
Or
A Measurement Extension
(M3015A or M3016A) is
connected to a standard
M3001A Measurement
Server (noninvasive meas-
urements only) and an
M3046A Monitor with
software Rev.E . No meas-
urements (CO
, 2nd Press/
2
Temp) from the Measure-
ment Extension are availa-
ble.
The M3046A Monitor
does not support some of
the measurements in a
Measurement Extension.
Use extensions only with
Extended Measurement
Servers (#C06 for
M3000A; #C06 and #C18
for M3001A).
,
2