5.1.
5.2.
5-1
PERFORMANCE CHECK
GENERAL INFORMATION
WARNING:
Before switching on, ensure that the instrument has been installed in accordance with the instructions
outlined in Section 2: of the Operating Manual Installation instructions.
This procedure is intended to:
check the instrument's specification
ATTENTION:
up time of 5 min.
GENERAL FUNCTIONAL TEST
Set the instrument to POWER ON
Disconnect any component or test cable/adapter from the. input sockets of the RCL meter
Check that display shows the following display reading:
® RCL AUTO
lf necessary, perform compensation of the zero-capacitance by Co TRIM according to chapter 3.4.3
of the operating manual.
Connect a precision resistor 0.1 % (e.g. 1 — 10 k&2) or a resistor of which the exact value is known
to the measuring input
Check the correct measuring result.