Changing Conductivity Limit Width - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Conductivity Alarm Window
2) Press the  (up) arrow key to
increase the distance between
the upper and lower alarm limits.
Press the  (down) arrow key to
decrease it.
Figure 34 – Changing Conductivity Limit Width
Warning! Always verify the conductivity and approximate pH of the dialysate solution
through independent means (e.g. using a conductivity meter or pH paper or meter, as
applicable) before initiating each dialysis treatment. Verify that the conductivity is
reasonably close to the theoretical conductivity value (TCD) and the pH is between 6.9 and
7.6. If they are not, do not initiate dialysis.
2008T Machine Operator's Manual P/N 490122 Rev N
Chapter 3—Setting Treatment Parameters
1) To access the
conductivity alarm window,
select Alarm Width.
3) Press CONFIRM
to save the new
alarm settings.
73