Chapter 3—Setting Treatment Parameters
1) The electrolyte profile of the
highlighted selection is
displayed in this column.
Note: If the machine is set up
for use with Citrasate a
'Citrate' meter box will be
displayed in the dialysate
constituent list.
2) Highlight the desired
selection by scrolling through
the choices using the  or 
(up or down) keys on the
keyboard. Acetate
formulations can also be
selected as a treatment option.
3) When the desired type is highlighted, press
the Enter key or CONFIRM key on the Touchpad
to save the selection and return to the "Dialysate"
screen.
Note: Be sure the acid/bicarbonate concentrates
connected to the machine match the type
selected from the Concentrate menu.
4) Enter the prescribed Base Na+
and Bicarbonate values as described
in "Entering a Treatment Parameter."
Figure 31 – Entering Concentrate Information
70