To set an SVS profile:
1) From the "Home"
screen, select the SVS
Profile button. The SVS
Profile window opens.
3) Select Start Na,
and enter the
starting maximum
+
Na
value using the
Keyboard.
4) Select Base Na
the Keyboard, enter the final Na
concentration value to be
reached at the end of the SVS
profile.
Figure 40 – Entering an SVS Program
The SVS timer is activated when the Tx Clock button is initially selected and confirmed to
start treatment. The SVS profile parameters can only be changed if the SVS is turned off
using the 'None' profile button.
2008T Machine Operator's Manual P/N 490122 Rev N
2) Select the prescribed profile by
selecting the corresponding button from
among the four options located in a row
along the top of the screen.
+
, and, using
+
5) Select the SVS-Time
button and, using the
Keyboard, enter the
running time for the SVS
profile.
Chapter 3—Setting Treatment Parameters
6) Press CONFIRM to
save the SVS settings.
The SVS Profile button
on the "Home" screen will
be colored green when
the selected SVS profile is
running (software version
2.34 or later).