Troubleshooting
Put Connectors in Port
The connectors must be in the
machine ports in order to start a
Rinse program

Put Lines On Shunt

An action has been initiated that
requires the dialyzer supply and
return lines to be on the shunt.
Put Red Con in Chemical
This is a cleaning/disinfectant
program prompt to the operator.

RAM Battery Failure

Electronic self-test failure.
RAM Code Corrupted 1
Electronic self-test failure.
RAM Code Corrupted 2
Repeated electronic self-test failure.

Recirc Interrupted

The Recirculate program has been
interrupted by an alarm condition.
Recirculating
Recirculation is in progress.
Recirculating Done
A prompt to the operator that the
recirculation process is done.

Recirculating Stopped

Recirculation has been stopped
because blood is sensed or the
dialyzer supply and return lines are on
shunt.
Release CONFIRM to
This message is a prompt for the
stop
operator to release the CONFIRM key
to stop priming the heparin line.
2008T Machine Operator's Manual P/N 490122 Rev N
Advisory
Connect the red (acid/acetate) and/or blue (bicarbonate) connectors
to the appropriate rinse ports.
Warning
To proceed with the selected operation, place dialyzer supply and
return lines on the shunt.
Advisory
Remove the red connector from the machine and place it into the
wand in the yellow chemical/disinfectant bottle.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Warning
1) Inspect the blood pump condition. Correct if required.
2) Reset the alarm and turn UF back on, if applicable.
3) If the alarm does not clear, take the machine out of service and
alert a qualified service technician.
Advisory
Advisory only. No action is required.
Advisory
Press RESET to clear the advisory message.
Warning
Inspect the configuration of the dialyzer supply and return lines and
extracorporeal blood circuit. Correct any irregularities. If the message
is not cleared, take the machine out of service and alert a qualified
service technician.
Dialog
Release the CONFIRM key.
Message