Fold-Down Touchpad; Touchpad Keys - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Fold-Down Touchpad

Figure 8 – Control Panel – Touchpad
The Touchpad is located directly below the Control Panel Keypad. It folds down to reveal a
touchpad which is used to move the on-screen cursor arrow. It also features two keys:
Table 4 – Touchpad Keys
Press ...
Touchpad
Note: The on-screen cursor arrow will disappear if not moved for five seconds. Move the
cursor using the Touchpad for it to reappear.
It also disappears when a value is entered but not yet confirmed. To display the cursor
again, press the Escape key to cancel the change or press the CONFIRM key to confirm the
change.
2008T Machine Operator's Manual P/N 490122 Rev N
To ...
Select a field highlighted by the on-screen cursor arrow.
Save a treatment parameter entry or confirm an action initiated on the
display screen. The CONFIRM key is a backup, safety feature designed
to prevent accidental changes to the intended treatment parameters.
Note: The CONFIRM key functions the same as the Enter key on the
Keyboard.
Void the current entry and return to previously entered parameter value
before CONFIRM is pressed.
Note: The Escape key functions the same as the Esc key on the
Keyboard.
Tap the Touchpad to select a field highlighted by the on-screen cursor
arrow
Tap the Touchpad again to confirm a change.
Chapter 1—Overview
29