212
Troubleshooting

Tx Clock On

The operator has selected the Empty
bibag button when the Tx Clock is
running. The bag cannot be emptied
using the Empty bibag button when
the Tx Clock is running.
Tx Clock Paused?
Blood is sensed in optical detector
while Tx clock is paused.
UF Goal greater than
Entry value for goal is out of range.
max. value

UF Goal Reached

This message is to alert the operator
that the preset ultrafiltration goal has
been reached.
UF Is On
An action has been initiated that
requires the UF to be off.
UF Profile Error
A UF profile calculation error has
been detected.
UF Pump Alarm
UF pump is not connected or is not
pulsing properly.

UF Pump Off

The UF pump has been off for ten
minutes.

UF Rate Error

A calculation error has been detected
UF Removed cleared
This temporary message is displayed
when the Tx clock is turned on the
first time after the New Tx key was
pressed. The UF removed has been
set to 0.
2008T Machine Operator's Manual P/N 490122 Rev N
Dialog
If the treatment is not yet completed and the bibag disposable must
Message
be changed, lift up on the bibag door, wait 30 seconds to relieve the
pressure, and remove the bag. Put a new bibag disposable on the
connector and close the door again to continue using the bibag
system for the treatment.
Warning
Start Tx clock
Dialog
Readjust UF Goal
Message
Warning
Press RESET to reset the alarm. The preset UF Goal has been
reached and the UF Rate will drop to the minimum UF Rate. If the
patient's prescribed UF Goal has not been reached, the operator
must take further action to comply with the prescribed treatment.
Advisory
To proceed with the selected operation, turn the UF pump off.
Alarm
Reset the UF parameters
Alarm
Press RESET to reset the alarm. If unable to clear the alarm, return
the blood to the patient if alarm occurs during treatment. Take
machine out of service and alert a qualified service technician.
Warning
To continue with the UF pump off, press RESET to clear the
message. This message will not be displayed again.
If you intend for the UF pump to be on, press the UF on/off key on
the control panel. The light above the key will be illuminated.
Alarm
Reset the UF parameters
Advisory
No action necessary.