Setting the Acid/Bicarbonate Alert

The Acid/Bicarbonate Alert option sounds an alarm when the fluid level in either of the
concentrate jugs has been drained to 20 percent of its original amount. In addition to the
alarm, a warning message such as LOW ACID WARNING , LOW BIC WARNING , or LOW
ACETATE WARNING will appear in the status box. This alert aids the operator in
maintaining adequate amounts of concentrate in the containers during treatment. Be sure to
set the new volume in this screen whenever the concentrate containers are refilled.
1) From the "Dialysate"
screen, select Acid/Bicarb
Alert. The "Acid/Bicarb
Alert" subscreen appears.
3) Select Bicarbonate. Using the keyboard,
enter the actual amount, in liters, of
concentrate in the bicarbonate jug.
Note: If the facility is
on a centralized acid
or bicarbonate supply
system, enter zero to
deactivate the alarm
for that concentrate.
6) Select Dialysate to exit the
"Acid/Bicarb Alert" subscreen.
The volume of each concentrate
will count down as it is used.
Figure 32 – Setting Acid & Bicarbonate Alerts
2008T Machine Operator's Manual P/N 490122 Rev N
2) Select Acid. Using the keyboard, enter the actual
amount, in liters, of concentrate in the acid jug.
5) Press CONFIRM to
save the change.
Chapter 3—Setting Treatment Parameters
4) Select Alert until an X
appears in the On checkbox.
71