Advertisement

Chapter 5—Disinfection and Maintenance
8. Chemical dwell follows after a delay. Remaining Time meter box counts down.
9. When the Remaining Time meter box reads 0:00, remove the red acid connector from the
disinfectant jug and insert it into the acid rinse port. The machine will automatically run
for about a minute to draw up the disinfectant left in the tubing.
Note: Visually confirm that disinfectant has been pulled into the machine.
10. Following the completion of the chemical disinfection cycle, "Press CONFIRM to exit"
will display in the Status Box.
11. Press CONFIRM to exit.
Warning! The mandatory rinse cycle must be completed and a test for residual disinfectant
must be performed prior to the next treatment.
148
2008T Machine Operator's Manual P/N 490122 Rev N