Failed To Print; Failed Sending Data To Actuator Board; Fill Program Alarm; Filling Program - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

192
Troubleshooting
Fail * Temp High Hard
Temperature test has failed.
Fail * Temp Low Hard
Fail * TMP High Soft
Transmembrane Pressure (TMP) test
has failed.
Fail * TMP Low Soft
Fail * TMP High Hard
Fail * TMP Low Hard
Fail * Ven High Soft
Venous Pressure test has failed.
Fail * Ven Low Soft
Fail * Ven High Hard
Fail * Ven Low Hard

Failed To Print

The CLiC device treatment report was
not printed.
** Failed Sending Data to
Functional to Actuator board
Actuator Board **
communication problem during
startup.
Fill for Diasafe Test
This message indicates the status of
the Diasafe test

Fill Program Alarm

A Fill program has occurred for one
minute while blood is sensed.

Filling Program

A Fill program is in progress.
2008T Machine Operator's Manual P/N 490122 Rev N
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Warning
Press the RESET key to clear the warning. Check the connections to
the printer and try to print again. If unable to print, contact a qualified
service technician.
Note: If you press the New TX key, the Crit-Line data for the
completed treatment will be lost.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Advisory
Advisory only. No action is required.
Alarm
Inspect for air in the system.
Correct as required.
If the warning is not cleared, return the blood to the patient if alarm
occurs during treatment. Take the machine out of service and alert a
qualified service technician.
Advisory
Advisory only. No action is required.

Advertisement

Table of Contents
loading

Table of Contents