Message
Purpose of Message
bibag: Cond Low,
continued

bibag: Cond Sensor Err

The bibag conductivity sensor is not
reading the correct conductivity. The
machine is in bypass mode.
167
Troubleshooting
Type
Action Required
(continued)
If conductivity alarm persists,
1) Turn off the dialysate flow by selecting and confirming the
Dialysate Flow on/off toggle-button on the "Dialysate" screen.
2) Leaving the bag on the connector, pull the blue bicarbonate
connector out of its rinse port.
3) Check for a clogged filter screen in the blue bicarbonate
connector handle. Clean if necessary with purified water. Verify
that the connector and filter assembly are tightly screwed
together with no air leak.
4) Insert the blue bicarbonate connector back into its rinse port.
5) Turn the dialysate flow back on by selecting and confirming the
Dialysate Flow on/off toggle-button on the "Dialysate" screen.
If the conductivity alarm still cannot be cleared, rinse back the
patient's blood and disconnect the patient from the machine. Run an
Acid Clean program followed by a complete rinse cycle. Test
machine operation.
If conductivity alarm still persists, take the machine out of service and
alert a qualified service technician.
Advisory
Insert the blue bicarbonate connector into an appropriate liquid
bicarbonate source
Or
To use the bibag disposable: If during treatment, rinse back the
patient's blood and disconnect the patient from the machine. Run a
Rinse program until message is cleared. If the alarm is not cleared,
take the machine out of service and alert a qualified service
technician.
ALARMS
TROUBLE-
SHOOTING
2008T Machine Operator's Manual P/N 490122 Rev N