Message
Purpose of Message
Select new Goal or Time
UF rate is out of range.
Select Program
This message is a prompt for the
operator to select a program.
Set Arterial Limits
This is a message to re-center the
arterial limits if necessary
Set Blood Flow to 300
This message is a prompt for the
operator to set the blood flow rate in
preparation for the Access Flow test
Set TMP Limits?
This is a message to re-center the
TMP limits if necessary
Set Venous Limits
This is a message to adjust the
venous limits if necessary
Short Power Down
The machine was switched off for 1 –
2 minutes and turned back on. Setup
values have not been set to default
values.
Single Needle On!
An action has been initiated that
requires the Single Needle option to
be off.
SN BP +5 V Error
+ 5 volts is outside the allowable
range
SN BP +12 V Error
+ 12 volts is outside the allowable
range
SN BP -12 V Error
- 12 volts is outside the allowable
range
SN BP +24 V Error
+ 24 volts is outside the allowable
range
SN BP Button Alarm
Key stuck or held in too long
207
Troubleshooting
Type
Action Required
Dialog
Enter a new UF Goal or reduce UF rate.
Message
Advisory
To proceed, select the desired program and press CONFIRM.
Advisory
Press and hold Reset for 1 second to re-center the limits.
Warning
To proceed, set Blood Flow to 300 and press CONFIRM.
Advisory
Rising TMP may indicate a leak in the balancing system and should
be investigated.
Press and hold RESET for 1 second to re-center the limits.
Advisory
Press and hold RESET for 1 second to adjust the limits. Changes in
venous pressure during the treatment should be investigated. See
"Ven. Pressure Alarm"
Advisory
Verify that the dialysis parameters are as desired.
Warning
To proceed with the selected operation, de-select the Single Needle
option.
Alarm
See message E.10
Alarm
See message E.07
Alarm
See message E.09
Alarm
See message E.08
Alarm
See message A.16
ALARMS
TROUBLE-
SHOOTING
2008T Machine Operator's Manual P/N 490122 Rev N