Must Be Alarm Free; Must Calibrate To Run; Must Clear Uf Removed; Negative Error Af Value - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

200
Troubleshooting

Must Be Alarm Free

A conductivity alarm exists when an
SVS program attempted to start.

Must Calibrate to Run

Electronic self-test failure.

Must Clear UF Removed

An action has been initiated that
requires the UF Removed to be
cleared to zero.
Must Run Test First
The Forced Test is required before
proceeding with UF or SVS.
Need Blood Sensed!
An action has been initiated that
requires that blood is sensed.
Neg. Access Flow value
This message is an advisory
message that the Access Flow test
result was a negative value. A
positive value is expected.

Negative error AF value

This message is an advisory
message that the Access Flow test
result was an erroneous value.
New Art Limits chosen
This message advises the operator
that a new set of arterial limits has
been set.
New features loaded,
Advisory message when uploading
Power Off, Replace
hardware key option
EEPROM
New TMP Limits chosen
This message confirms that a new set
of TMP limits have been set.
New Ven Limits chosen
New venous alarm limits are set
No Air Detector Alarm
The Prime key has been pressed. A
level detector alarm must exist for this
function to occur
2008T Machine Operator's Manual P/N 490122 Rev N
Advisory
The machine must be alarm free in order to run an SVS program.
Correct the alarm before starting the SVS program.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Warning
To proceed with the selected operation, set the UF Removed
treatment button to zero.
Warning
To proceed with the selected operation, run the Pressure and Alarm
tests. For more information, see Chapter 2 "Testing the 2008T
Hemodialysis Machine."
Verify venous bloodline is in the Optical Detector.
Dialog
Message
Warning
This can occur if the bloodlines were initially connected in the
reversed position. Press CONFIRM to clear the message. Check that
the bloodlines are properly connected and repeat the Access Flow
test.
Warning
Press CONFIRM to clear the message. Check that the bloodlines are
properly connected and repeat the Access Flow test. Press
CONFIRM to clear the message.
Advisory
Advisory only. No action is required.
Advisory
Put the original calibration EEPROM in IC 20 and restart the
machine.
Advisory
Advisory only. No action is required.
Advisory
Advisory only. No action is required
Advisory
If the venous chamber has fluid detected, the prime function will not
occur. Press RESET to start the blood pump. If a level detector
alarm occurs, then press the Prime key.