Performance Limits; Alarm Preset Values For Adult - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Appendix C
Blood Pressure Module
Technique
Cuff Deflation
Cuff Inflation
Interval Settings
Cuff Pressure Range
Initial Cuff Inflation
Systolic Determination Range
MAP Determination Range
Diastolic Determination Range
Pulse Rate Determination Range
Cuff Inflation Rate
Determination Time Normal
Overpressure Cut Off
Transducer Drift
Leakage Rate (Max)
Pressure Rate Offset
Systolic
MAP
Diastolic
Pulse
Inflation Pressure
270
Measures systolic, diastolic pressures, and heart rate (pulse
rate) using oscillometric method. MAP measured.
Interactive computer controlled. Determination for adults
requires approximately 25-30 seconds depending on starting
point, heart rate and motion artifact.
Typically 5 - 10 seconds from 0-250 mmHg
Interval times: 5 - 60 minutes in increments of 5 minutes
Clock Time: 5, 10, 15, 20, 30, 60 minutes

Performance Limits

10-325 mmHg
180 mmHg or adjusted by host
60-260 mmHg
35-220 mmHg
30-200 mmHg
25-30 seconds
3 mmHg/min in 3 minutes

Alarm Preset Values for Adult

(Internal alarm values preset to provide alarm limits in the event
individual values are not entered)
2008T Machine Operator's Manual P/N 490122 Rev N
40-180 BPM
5 seconds
325 mmHg
Auto Zeroing
Auto Zeroing
200/80
120/70
110/50
120/50
Auto