Chapter 1—Overview
Press ...
To ...
(New Treatment) Erase the current treatment information and move the
summary information to the previous record in the "Trends" screen.
Press the CONFIRM key on the Touchpad or Enter on the Keyboard to
complete the action. To cancel, press the Escape key on the Touchpad or the
Esc key on the Keyboard.
Reset the machine after an alarm.
Press and hold for two seconds to spread the alarm window by 300 mm Hg for
arterial and venous pressures and fully open the transmembrane (TMP)
pressure window for 30 seconds. The light above the RESET key will not be on.
During a blood leak alarm, press and hold for three seconds to override the
alarm and keep the blood pump running for three minutes. The light above the
RESET key will be on during an override.
Start an unscheduled, manual blood pressure measurement when the cuff is
deflated, or instantly deflate the inflated blood pressure cuff.
Turn the ultrafiltration pump on or off. During ultrafiltration, the green light is
illuminated. This light will flash when ultrafiltration is interrupted.
Prime the extracorporeal blood circuit. Pressing Prime will keep the blood
pump running when air is sensed in the venous blood chamber and an air
detector alarm is present (as is the case during initial set up when the blood
circuit tubing is empty). The pump will run for:
Two minutes, or
Until an adequate fluid level is detected by the ultrasonic sensors in the level
detector module, or
Until the volume set in Service Mode is reached.
Note: Pressing any control panel key (except for the Power key) while in CDX Mode (if
available) will switch the machine back to Dialysis Mode. See Appendix B on page 233 for
more information on CDX Mode.
24
Note: The RESET key is only used to reset alarms; it does not reset
or cancel changes to a parameter.
Note: Certain versions of the blood pressure module require a 30
second delay between blood pressure measurements.
Note: Pressing the Stat/Deflate key while in CDX mode will only exit
CDX mode. The operator must then press the Stat/Deflate key
again in order to take a blood pressure measurement.
Note: When the UF pump is turned off, there is no "minimum"
ultrafiltration occurring.
2008T Machine Operator's Manual P/N 490122 Rev N