Machine Specifications

Dimensions

Floor space
Height
Total weight
Operating conditions
Storage conditions
Electrical
Power Supply—Main
Power Consumption
Fuses
External Connections
Heat Dissipation to Room
Electromagnetic compatibility
Electrical safety (UL 60601-1)
Protection against electric shock
Leakage currents
Water
Ingress Rating
Back Flow Prevention
Water Pressure
Water Temperature
264
Approximately 54 cm wide by 63 cm deep
149 cm
Approximately 90 kg
60 – 100 °F (15.5 – 38 °C)
Relative Humidity 10% to 90%, non-condensing
Room temp, 6 months, Do not freeze
Relative Humidity 10% to 90%, non-condensing
Single phase AC 117 V ±10% 60 Hz ±3 Hz must be
connected to a circuit which is equipped with a hospital
grade receptacle and is protected by circuit breaker and
ground fault interrupter (GFI). Resistance from chassis to
ground must be < 0.2 ohm.
Does not exceed 12.5 amps
6.3 amp medium blow fuse, 2 each
16 amp double pole rocker switch circuit breaker for heater
Isolated RS232 and Ethernet ports; leakage current
isolation per UL 60601-1 between the machine and external
computer
600 to 700 BTU/hr
See EMC Declaration on page 271
Type: Safety class I Degree: Type B
Type CF: Only BPM Blood Pressure Cuff
According to UL 60601-1
IPX1 (Vertical drip-proof protection)
Integral back flow prevention provided by external vent to
atmosphere in water inlet circuit.
Min 20 psi; max 105 psi
Min 10 °C; max 25 °C
2008T Machine Operator's Manual P/N 490122 Rev N