Setting Auto Heat Disinfect Schedule - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Appendix C
3) Select Start
Prg and, using
the keyboard,
enter the time at
which the Heat
Disinfect should
automatically
begin.
Note: The clock
is a 24 hour
clock, so 0:00 is
midnight and
13:00 is 1:00
PM.
Figure 82 – Setting Auto Heat Disinfect Schedule
7. Exit Service Mode and go to the "Start-Up" screen after powering on.
8. Ensure that both dialysate lines are on the shunt.
9. Place both concentrate connectors in their respective ports. The Auto Heat Disinfect
program will automatically run at the selected time(s).
Warning! Do not run a Rinse program at the end of the treatment day before allowing the
Auto Heat Disinfect program to run. Doing so will not allow the RO (permeate or product)
water to flow into the machine during the Auto Heat Disinfect program. Leave the machine
on in the "Select Program" screen with the concentrate connectors in their ports at the end of
the treatment day. The Heat Disinfect will then run automatically.
Note: The dialysate lines must be on the shunt and both concentrate connectors must be in
their respective ports for the Auto Heat Disinfect program to run.
Note: At the end of the treatment day, perform an Acid Clean and leave the machine on the
"Select Program" screen.
254
6) Press CONFIRM.
2008T Machine Operator's Manual P/N 490122 Rev N
4) Select the desired
day(s) of the week
button(s) on which the
Auto Heat Disinfect
should run. The blue
X will change from Off
to On, indicating the
Heat Disinfect will run
on that day at the
specified time (set in
the Start Prg button).
5) Select Pre-Rinse
Time button and enter
the desired time for a
rinse that runs before
the Auto Heat Disinfect
program.
Note: The heater will be
off during the pre-rinse.

Advertisement