Troubleshooting

Take Lines Off Shunt

An action has been initiated that
requires the dialyzer supply and
return lines to be off the shunt.
Target Kt/V has been set
The operator has attempted to set the
to min.
Target Kt/V to less than the minimum
allowed
Temp DAC Error
The DAC (Digital/Analog conversion)
for the temperature trim function is
outside of its limits.
Temp Over 95 Degrees
Electronic self-test failure.
Temp Sensors
When the temperature trim function
unmatched
needs to change DAC by > 1 C, the
pre and post sensors are verified
against one another. This message
occurs if the two temperature sensors
are more than 0.5 C different.
Temp Control not
This message is displayed on the sign
calibrated
on screen if the temperature sensors
were not matched when they were
verified against one another.
Temperature greater than
Entered Temperature value is higher
max. value
than allowed.
Temperature has been
The operator has attempted to set a
set to min.
Temperature lower than allowed.
Test Complete
All selected self-tests passed.
Test Failed
The Alarm and/or PHT Sections of the
automated Test Sequence have failed
2008T Machine Operator's Manual P/N 490122 Rev N
Warning
To proceed with the selected operation, dialyzer supply and return
lines must be off the shunt. Connect lines to the dialyzer.
Dialog
The machine has set the target Kt/V to the lowest allowed target.
Message
Verify that the value is acceptable.
Warning
Press RESET key to reset alarm
The temperature trim function will be disabled until the temperature
sensors are recalibrated.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Warning
Press RESET key to reset alarm
The temperature trim function will be disabled until the machine is
turned off and back on.
Advisory
Calibrate the temperature control.
Dialog
The temperature will be set to the highest allowed level. Press
Message
CONFIRM to clear message and accept the maximum allowed value.
Verify that the value is acceptable or enter new value.
Dialog
The temperature will be set to the lowest allowed level.
Message
Advisory
Advisory only. No action required.
Test
Reset the alarm. Check the setup to see if the alarm can be
Alarm
corrected and then retest. If the machine fails, turn machine power
Off and back On. If alarm is still not cleared, take the machine out of
service and alert a qualified service technician.