Proportional Mixing System Data - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Appendix C
Proportional Mixing System
Acid
Acetate
Adjustment Range
Bicarbonate
Adjustment Range
Monitoring Conductivity
Optional bibag Dry Bicarbonate
Range of Display
Dialysate Heating
Nominal Value of Temperature
Temperature Display
266
Volumetric, selectable:
1:34
1:44
1:35.83
1:35.1
Note: Citrasate is for use with 1:44 concentrates only.
1:34
+
130 to 155 mEq/L Na
Volumetric, selected with associated acid ratio
1:27.46
1:19.13
1:25.16
1:27.6
20 to 40 mEq/L Bicarbonate (post-reaction, after mixing with the
acid and purified water).
Average Accuracy: ± 1.5%
Method: Temperature compensated electronic conductivity
meter with adjustable alarm limits.
Temperature compensated conductivity display with
automatically set alarm windows ±0.5 mS/cm around calculated
conductivity*. User can adjust an additional ±0.5 mS/cm within
this range.
Temperature compensated conductivity display with
automatically set alarm windows ±0.5 mS/cm around calculated
conductivity*, limited to ±0.4 mEq/L @ 24 mEq/L bicarbonate or
less.
With alarm window set at ±0.5 mS/cm:
User can move alarm window up or down an additional:
±0.2 mS/cm @ 40 mEq/L
±0.1 mS/cm @ 35 mEq/L
no adjustment @ 29 mEq/L
*Conductivity is based on the concentrates' compositional data
entered in the Dialysate screen at the standard temperature of
25 C.
10.0 to 17.0 mS/cm. At 25 °C. Alarm limits will not go below 12.5
or above 16.0 mS/cm.
35 to 39 °C Accuracy: ± 0.3 °C (measuring accuracy under
calibration conditions for a dialysate flow of 500 ml/min)
(selectable in 0.1 °C steps)
Range 35 to 39 °C with alarm limit window automatically
adjusted to 2 °C above and below set point. Alarm window will
not go below 30° or above 41 °C. Heater 1.3 kW, electronically
controlled.
2008T Machine Operator's Manual P/N 490122 Rev N