Troubleshooting

Conductivity out of range

When attempting to start an OLC test,
certain conditions are necessary,
including the condition that
conductivity for both the inlet and
outlet sensors be in range.

CONFIRM Concentrate

This message will be displayed if the
user needs to confirm the concentrate
selected for use.

Connector(s) Out Of Port

An action has been initiated that
requires the Acid/Bicarbonate
Connectors to be in their rinse ports.
Cooling Down
The machine is cooling down from a
heat disinfect.
** Cover is Open **
The dialysate shunt door is open.
Crit-Line disabled. Must
The CLiC device must be verified
verify Crit-Line.
before it is available for use during
treatment.
2008T Machine Operator's Manual P/N 490122 Rev N
Advisory
Wait until the conductivity is stable and start the OLC test again. If
the message repeats, do not use OLC until the conductivity sensors
have been recalibrated.
Advisory
Press CONFIRM or change the concentrate selection and then press
CONFIRM.
Warning
Insert the concentrate connectors into their proper rinse ports.
Advisory
Advisory message only. No action is required.
Advisory
To proceed with the selected operation, close the shunt door.
If this message is displayed during the treatment:
Advisory
message
The CLiC device is not monitoring the patient. The CLiC
device cannot be used until the next treatment.
If the message is displayed before the treatment starts:
Place the CLiC device on the verification filter, which is
attached to its USB cable.
Wait up to one minute for the message to clear. If the
message clears, the CLiC device is verified.
If verification fails, select the "Test & Options" screen-
button and select the Verify Crit-Line button to initiate the
verification process manually.
If the message is still not cleared, the CLiC device cannot be used
for the dialysis treatment. Disconnect the CLiC device cable from
the USB port and contact a qualified service technician.