Message
Purpose of Message
BTM test underway
The OLC test may not be started
when a BTM recirculation test is
underway.
BVM Failed
The BVM module has failed
BVM No Communication
The BVM module has lost
communication with the 2008T
system.
CDX blocked due to
The operator has attempted to enter
alarm
CDX mode during an alarm.
CDX Not Active
The optional CDX mode is not
enabled.
CDX Not Installed
The optional CDX mode is enabled
but the CDX system PC is not
installed in the machine.
Check cable or press 'Y'
The 2008T hemodialysis machine has
to disable Crit-Line
not received data from the CLiC
device. The Status Box displays the
warning message "Crit-Line: No
Comm"
Chem not Connected?
The red acid connector is still
connected to the red rinse port
181
Troubleshooting
Type
Action Required
Advisory
Wait for the BTM recirculation test to complete before beginning the
OLC test.
Alarm
Press RESET to clear the message. BVM will no longer pass
information to the 2008T monitor. Turn the power off and back on. If
the alarm is not cleared, refer to a qualified service technician.
Alarm
Press RESET to clear the message. BVM will no longer pass
information to the 2008T monitor until power has been turned off and
back on. If alarm is not cleared, return the blood to the patient if
alarm occurs during treatment. Take the machine out of service and
alert a qualified service technician.
Advisory
The machine must be alarm free in order to access the CDX system.
Correct the alarm before accessing the CDX system.
Dialog
Advisory message only. No action is required. Alert a qualified
Message
service technician if CDX mode needs to be enabled.
Dialog
Advisory message only. No action is required. Alert a qualified
Message
service technician if the CDX system needs to be installed.
Advisory
Check the CLiC device's USB cable:
message
If the Status Box warning message "Crit-Line: No Comm" is not
cleared, the CLiC device cannot be used for the dialysis treatment.
Warning
Connect the red (acid) connector into its correct configuration for the
operation selected.
ALARMS
TROUBLE-
SHOOTING
Make sure the cable is securely connected to the 2008T
hemodialysis machine's USB port.
Wait up to one minute for the message to clear.
Disconnect the CLiC device cable from the USB port.
Press the 'Y' key on the Keyboard to disable the CLiC device.
Contact a qualified service technician.