The Cdx System (Optional) - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Appendix B

The CDX System (Optional)

The Clinical Data eXchange (CDX) system is an independent PC that is built into the 2008T
hemodialysis machine. The CDX PC can run software programs that are displayed on the
2008T display screen and accessed with the Keyboard, Touchscreen, and Touchpad like any
other personal computer (PC). Pressing the blue CDX key on the keyboard (see Figure 71)
will switch between Dialysis Mode and CDX Mode.
Figure 71 – The Keyboard
The 2008T hemodialysis machine will automatically switch from CDX Mode to Dialysis
Mode:
During hemodialysis alarms
By pressing any control panel key except for the Power key (see Figure 72 on the next
page)
If no input (keyboard, touchpad, etc.) is received after two minutes when blood is sensed
and the Tx Clock is running
Note: CDX mode is unavailable during an alarm when blood is sensed and the Tx Clock is
running.
2008T Machine Operator's Manual P/N 490122 Rev N
Appendix B
The CDX key
233