Chapter 5—Disinfection and Maintenance
7. After the heat disinfection is complete, if the machine is not configured to automatically
turn off at the completion of the cycle, press CONFIRM to exit when prompted.
Note: The drain line is subjected to a lower temperature and shorter heat cycle than the rest
of the machine. If you are unable to completely clean biofilm from the drain line, select the
"Extended Pre-rinse" option in Service Mode. If necessary, replace the drain line.
Note: Cooling time can be shortened by running the Rinse program, which will flush the
machine with 37°C water. Do not cool the machine with the Rinse program unless the
machine will be used immediately afterwards.
150
2008T Machine Operator's Manual P/N 490122 Rev N