Auto Flow Service Mode Selections

Figure 80 – Auto Flow Service Mode Screen
Table 39 – Auto Flow and Idle Mode Features
Button
2008T Machine Operator's Manual P/N 490122 Rev N
Function
The Auto Flow Minimum option selects the dialysate flow rate during
dialysis when Auto Flow is selected:
300 – 800 ml/min or
500 – 800 ml/min (default selection)
There are four available choices for Auto Flow in dialysis:
Both—The operator can scroll to either 1.5x or 2x (default
selection).
1.5X—The operator only has a choice of 1.5x (plus the normal flow
rates).
2.0X—The operator only has a choice of 2x (plus the normal flow
rates).
None—The operator cannot choose any Auto Flow options.
See the blood flow rate table on page 265 for more information.
The Default Dial Flow toggle-button is used to set the default
dialysate flow rate the machine runs after exiting Idle Mode, either
500 or 800 ml/min (if 1.5x or 2x Auto Flow was not set). The user is
then free to set the prescription.