CO- 3.4.0 [CONDUCTIVITY] WITHIN 0.4 OF TCD / VERIFY CONDUCTIVITY ALARM LIMITS
What color is the Home screen's [Conductivity] window now? TWO (2) possible scenarios:
1)
IF (and ONLY if) pale yellow / white: See procedure number
2) IF RED: See parts a THROUGH e below:
a) At the bottom of the screen, press the 'Dialysate'
tab.
b) Figure right, if necessary adjust the Limits so that
the 'Actual' Conductivity is CENTERED
between them.
c) Press 'Enter'!
d) Allow up to two (2) additional minutes for the
'Actual' Conductivity window to turn white!
e) See procedure number
CO- 3.5.0 VERIFY MEASURED CONDUCTIVITY
a) Allow up to two (2) additional minutes OR until the external flow indicator's 'bob' starts to move up and
down!
b) Based on the meter's reading versus the machine's [Conductivity] window. TWO (2) possible
scenarios:
1)
IF (and ONLY if) the meter and the [Conductivity] window are within 0.15 of each other:
page 410
Proceed to
2) IF the meter and the [Conductivity] window are NOT within 0.15 of each other! Proceed to
page 383
, procedure number
Figure 57 – Hydraulics Rear View
CO- 3.5.0
(page 374).
, procedure number
CO-
CO-
6.0.0.
2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 3.5.0
(page 374).
9.8.0.
Page 374