No Chemical Intake; No Na Selected; No Water; Olc Test Failed - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message

No Chemical Intake

During the main program of chemical
rinse or chemical dwell, the machine
cannot get any chemical in the acid
connector.
+
No Na
Selected
This is a prompt to the operator that a
Start Na
set.
No Program Selected
This is a prompt to the operator that a
Profile was not selected.
No SVS Time Selected
This is a prompt to the operator that
the SVS Time has not been set.

No Water

A water inlet valve alarm has
occurred. The machine is not
receiving enough water.
OLC allowed in X minutes
There is a minimum waiting period
necessary between OLC tests
OLC steps not calculated
In order to do the OLC test, the
machine must calculate the pump
steps necessary to raise and lower
the conductivity for the test. This
cannot be done until stable
conductivity has been achieved.
OLC Test Cancelled!
User has cancelled OLC self test or a
condition occurred during the test
causing it to cancel.

OLC Test Failed

OLC self test failed.
OLC Test Passed
OLC self test passed
Online Clearance Self-
Machine is running an OLC self-test
test
201
Troubleshooting
+
value for SVS has not been
ALARMS
TROUBLE-
SHOOTING
Type
Action Required
Alarm
Retry chemical rinse and if problem persist, remove machine from
service and alert a qualified service technician.
Advisory
To proceed with the SVS operation, set a value for Start Na
"SVS" subscreen.
Advisory
To proceed with an SVS operation, select an SVS Profile from the
"SVS" subscreen.
Advisory
To proceed with the SVS operation, set the SVS Time button in the
"SVS" subscreen.
Warning
Inspect the treated water source supplying the machine. Correct as
required. If the alarm does not clear, return the blood to the patient if
alarm occurs during treatment. Take the machine out of service and
alert a qualified service technician.
Advisory
Wait the indicated time and start the OLC test again.
Advisory
Wait a couple of minutes after the conductivity is stable and start the
OLC test again.
Advisory
Advisory only no action required
Warning
Restart machine to rerun the OLC self test.
Advisory
Advisory only no action required
Advisory
Advisory only no action required
2008T Machine Operator's Manual P/N 490122 Rev N
+