Fail Arterial High/Low Soft/Hard; Fail Blood Leak; Fail (Get Neg Tmp)/Fail (Get Pos Tmp); Fail (Remove Air) - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message
Fail * Arterial High Soft
Arterial Pressure test has failed.
Fail * Arterial Low Soft
Fail * Arterial High Hard
Fail * Arterial Low Hard
Fail * Blood Leak 1
Blood Leak test has failed.
Fail * Blood Leak 2
Fail * Cond High Soft
Conductivity test has failed.
Fail * Cond Low Soft
Fail * (Get Neg TMP)
Get Neg TMP test has failed.
Fail * (Get Pos TMP)
Get Pos TMP test has failed.
Fail * Neg Flow On
Negative flow on pressure holding
test failed
Fail * Neg Stabilize
Negative flow stabilize test failed.
Fail * Optical Detect
Optical Detector test has failed.
Fail * Pos Flow Off
Positive flow off pressure holding test
failed
Fail * Pos Stabilize
Positive flow stabilize test failed.
Fail * (Remove Air)
Remove air test failed.
Fail * Temp High Soft
Temperature test has failed.
Fail * Temp Low Soft
191
Troubleshooting
Type
Action Required
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Verify absence of air bubbles in flow indicator. Rerun test. If failure
Message
repeats, remove machine from service and alert a qualified service
technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Check UF pump. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate.
ALARMS
TROUBLE-
SHOOTING
2008T Machine Operator's Manual P/N 490122 Rev N