Troubleshooting; Water/Dialysate Alarms/Other Alarms; Status Box Message; Purpose Of Message - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Water/Dialysate Alarms
During a water/dialysate alarm (temperature or conductivity), the blood system continues to
operate, but the dialysate fluid is internally bypassed around the dialyzer. This can be
verified by visually inspecting the flow meter in the dialysate supply line. During bypass, the
float will remain stationary at the bottom of the sight glass.
A flow alarm will not cause the machine to go into bypass. Water/dialysate alarms are self-
resetting when the alarm condition is corrected. Temperature and conductivity alarms do not
occur during the pure UF mode of Sequential dialysis when there is no dialysate flow.
Other Alarms
Other alarms may be associated with other components, such as the Heparin or UF pumps,
BPM, BVM, BTM, etc.

Troubleshooting

All status messages (operational alarms, warnings, dialogues, and advisories) are displayed
on the control panel screen. These messages are generated due to conditions and events that
occur in the machine during operation. These messages will reset when the condition causing
the message is corrected. In some cases, the operator must reset them.
The table following this section is indexed by Status Box message. The table consists of four
columns:

Status Box Message

Message Purpose
Message Type
Action Required
Status Box Message
The Status Box Message column identifies the message as it appears in the Status Box or in
the Dialogue Box of the display screen.

Purpose of Message

The Purpose of Message column is a brief explanation of the Status Box message or the
condition that generated it.
2008T Machine Operator's Manual P/N 490122 Rev N
Chapter 6—Alarms and Troubleshooting
157