To Set The 2008T Hemodialysis Machine For Sequential Dialysis - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Chapter 3—Setting Treatment Parameters
To set the 2008T hemodialysis machine for sequential
dialysis:
Figure 38 – Setting Sequential Dialysis
After sixty minutes of pure ultrafiltration, an alarm sounds and the warning message, " 60
MINUTES FLOW OFF ," appears in the Status box. The operator has the option of
continuing pure ultrafiltration or starting dialysis. This alarm occurs only once.
To continue pure ultrafiltration, press the RESET key on the control panel keypad. This
will silence the alarm and terminate the warning message.
To start the treatment, select the Dialysate Flow button in the "Home" screen, set it to the
prescribed rate using the Keyboard, and press CONFIRM.
82
3) Use the 
arrow key to scroll
down to OFF and
then press the 
arrow key to
'SEQ'.
2008T Machine Operator's Manual P/N 490122 Rev N
1) Set UF treatment
parameters. Set the UF
Time to include the
combined treatment and
pure UF times.
2) Select Dialysate
Flow to highlight it.
4) Press CONFIRM This
will deactivate the "Flow
Off" warning for 60
minutes. The Dialysate
Flow button will display
'SEQ'.

Advertisement

Table of Contents
loading

Table of Contents