180
Troubleshooting
BPM: Oscil Wave Check
The diastolic blood pressure reading
is close to or greater than the systolic
pressure reading.
BPM: Pulse > 100
Patient's heart rate is above 100
beats per minute.
BPM: Pulse Amp Unif
The amplitude of the pressure pulses
is inconsistent with an accurate blood
pressure profile.
BPM: Pulse High
The latest pulse reading is above the
Upper Pulse alarm limit.
BPM: Pulse Low
The latest pulse reading is below the
Lower Pulse alarm limit.
BPM: Pump On > 30 sec
The pump that inflates the cuff has
been running longer than 30 seconds.
BPM: Systolic High
The systolic blood pressure reading is
above the set Upper Systolic alarm
limit.
BPM: Systolic Low
The systolic blood pressure reading is
below the set Lower Systolic alarm
limit.
BPM: Weak Pulse
The pulse pressure is too weak to
register an accurate measurement.
BPM: Zero Pressure
No pressure is detected by the blood
pressure module.
2008T Machine Operator's Manual P/N 490122 Rev N
Blood
Indicates a hardware malfunction. If problem persists, remove cuff
Pressure
and alert a qualified service technician.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant. May also indicates a hardware malfunction.
Alarm
Blood
Check pressure cuff for proper fit and alignment. Observe the patient
Pressure
for physiologic changes. Treat as patient as symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Cuff is not inflating. Check for loose tubing connections or a leak in
Pressure
the cuff.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Check the cuff for proper fit and inflation. Observe the patient for
Pressure
physiologic changes. Treat patient as symptoms warrant.
Alarm
Blood
Check for a loose connection in the inflation system. Correct as
Pressure
necessary. If no leak found, turn the power off, then back on. If
Alarm
problem persists, alert a qualified service technician.