Troubleshooting

Diasafe Test Passed

This message advises the operator of
the status of the Diasafe self test.

Diasafe Test Recovery

This message advises the operator of
the status of the Diasafe self test.
E.01 (Arterial or SN Blood
EPROM CRC error
Pump Message)
E.02 (Arterial or SN
Flash ROM CRC error
Blood Pump Message)
E.03 (Arterial or SN
RAM check error
Blood Pump Message)
E.04 (Arterial or SN
Reference Voltage error
Blood Pump Message)
E.05 (Arterial or SN
EEPROM error
Blood Pump Message)
E.06 (Arterial or SN
Watchdog timeout
Blood Pump Message)
E.07 (Arterial or SN
+ 12 volts is outside the allowable
Blood Pump Message)
range
E.08 (Arterial or SN
+ 24 volts is outside the allowable
Blood Pump Message)
range
E.09 (Arterial or SN
- 12 volts is outside the allowable
Blood Pump Message)
range
E.10 (Arterial or SN
+ 5 volts is outside the allowable
Blood Pump Message)
range
2008T Machine Operator's Manual P/N 490122 Rev N
Advisory
Press the RESET key to clear the message
Advisory
Advisory message only. No action is required.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.