Press Confirm To Exit/Load/Start; Press Escape To Cancel Rinse; Press Escape To Stop (Item); Pressure Test Failed - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message
Press CONFIRM to exit
This message is a prompt for the
operator to press the CONFIRM key
to exit the Rinse program.
Press CONFIRM to Load
This message is a prompt for the
operator to press the CONFIRM key
to load the heparin syringe.
Press CONFIRM to Start
This message is a prompt for the
operator to press the CONFIRM key
to start the program.
Press Confirm to switch
The operator has selected the O
to Blood Pressure graph.
graph field on the "Crit-Line" screen;
Press Escape to return to
the machine is prompting the operator
O2 graph
to choose between displaying the BP
graph or the O
Press Confirm to switch
The operator has selected the BP
to O2 graph. Press
graph field on the "Crit-Line" screen;
Escape to return to BP
the machine is prompting the operator
graph.
to choose between displaying the O
Sat graph or the BP graph.
Press ESCAPE to cancel
This message is a prompt for the
rinse
operator to press the Escape key to
cancel the Rinse program.
Press ESCAPE To Stop
This message is a prompt for the
[Item]
operator to press the Escape key to
stop loading the heparin syringe or
the Rinse program.

Pressure Test Failed

The pressure test section (PHT) of
the automated Test Sequence has
failed.
Priming
The operator has pressed the Prime
key and initiated the priming function.
203
Troubleshooting
Type
Advisory
Dialog
Message
Advisory
Sat
Advisory
2
message
Sat graph.
2
Advisory
message
2
Advisory
Dialog
Message
Alarm
Advisory
ALARMS
TROUBLE-
SHOOTING
Action Required
To proceed with the selected operation, press CONFIRM.
To proceed with the selected operation, press CONFIRM.
To proceed with the selected operation, press CONFIRM.
Press the CONFIRM key to switch to the Blood Pressure graph.
Or
Press the Escape key to return to the Oxygen Saturation graph.
Press the CONFIRM key to switch to the Oxygen Saturation graph.
Or
Press the Escape key to return to the Blood Pressure graph.
To proceed with the selected operation, press Escape then press
CONFIRM.
To proceed with the selected operation, press Escape then press
CONFIRM.
Reset the alarm and repeat the test. If the failure message is
repeated on retest, take the machine out of service and alert a
qualified service technician.
Advisory only. No action is required.
2008T Machine Operator's Manual P/N 490122 Rev N