Power Resumption Procedure - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Power Resumption Procedure

1. Press the POWER key to restore power to the machine. The screen displays the "Select
Program" screen with the message, "POWER FAIL RECOVERY."
2. Select the Dialysis button to enter the "Dialysate" screen.
3. In the "Dialysate" screen, check the conductivity settings (Na
type) and alarm limits. Verify that the dialysate concentration settings are correct. If not,
reset them.
4. Press CONFIRM to save the dialysate settings.
5. Select the Home button to display the "Home" screen.
6. Press the RESET key to reset any alarms. Conductivity and temperature alarms will reset
automatically when acceptable limits are reached—usually in about 3–5 minutes. If the
dialysate lines were disconnected, reconnect the dialysate lines when conductivity and
temperature return to their prescribed limits.
7. Insert venous line in venous clamp and optical detector.
8. If not still connected, reconnect the patient per unit policy. If you are performing single-
needle dialysis, re-insert the pump segment into the single-needle pump.
9. Press the blood pump Start/Stop key to restart the blood pump. Reset the blood pump to
the prescribed flow rate.
10. Select the Tx Clock button to resume dialysis and then press CONFIRM.
11. If the heparin pump or the Single-Needle option were active prior to the power failure,
reinitiate these functions upon power resumption.
12. The SVS program parameters are stored during a power failure. Restart the SVS Profile
program by selecting and confirming the desired SVS Profile. Adjust the SVS-Time if
necessary.
13. The UF treatment parameters are also saved during a power failure. Check all parameters
(UF Goal, UF Time, UF Rate, UF-Removed) for correct settings and adjust if necessary.
2008T Machine Operator's Manual P/N 490122 Rev N
Chapter 4—Monitoring the Treatment
+
, Bicarbonate, concentrate
131