Rel. Blood Volume Low; Remove Usb Device 1/2; Reset To Adjust Tmp; Reset Treatment? Confirm Or Escape - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message

Rel. Blood Volume Low

The BVM module has reported a
relative blood volume below the lower
limit.
Or
The current blood volume percentage
or hematocrit has dropped below the
BV Alert Level set on the "Crit-Line"
screen. The UF pump has been
turned off.
Remove USB Device 1
A USB device has been plugged into
the USB port on the right side of the
display screen.
Remove USB Device 2
A powered USB device has been
plugged into the USB port on the back
of the machine.

RESET to adjust TMP

The TMP exceeded the hard alarm
limits. The operator is given the
option to relieve the pressure to bring
the TMP within limits.
Reset Treatment?
The New Tx key has been pressed
CONFIRM or Escape
205
Troubleshooting
Type
Alarm
Alarm
Warning
Warning! Adjusting the
TMP repeatedly will
decrease the UF removed
from the patient.
Advisory
ALARMS
TROUBLE-
SHOOTING
Action Required
Alert
Press RESET to clear the message if using the BVM (Blood Volume
Monitor) module.
Assess the patient for any changes in physiologic state.
If using the CLiC device, the alert can be cleared by setting the BV
Alert Level button on the "Crit-Line" screen to OFF.
To turn the BV Alert level OFF, select the BV Alert Level
button, enter "0" (zero) on the keyboard, and press the
CONFIRM key.
After clearing the alert, the UF pump will automatically turn
back on to the UF Rate set on the "Home" screen.
Note: At the direction of a physician, a new BV Alert Level
can be selected.
To change the BV Alert Level, select the BV Alert Level
button, enter the desired value using the keyboard and
press the CONFIRM key.
Remove the USB device.
Remove the USB device. Turn machine power off and back on. If
alarm is not cleared, return the blood to the patient if alarm occurs
during treatment. Take the machine out of service and alert a
qualified service technician
Press the RESET key to reset the TMP alarm limits. Press and hold
the RESET key to re-center the limits.
Warning! Rising TMP may indicate a leak in the balancing
system and should be investigated.
To erase the current treatment information and move it to the
"Trends" screen when initiating a new treatment, press CONFIRM.
To cancel, press Escape.