Chapter 6—Alarms and Troubleshooting
Figure 68 – Control Panel Features for Blood Alarms
Table 33 – The Control Panel Keys Used During Alarms
Press...
156
Press to
silence
alarm
To...
Silence an alarm for two minutes or until another alarm event occurs.
The red Mute light illuminates.
Press to reset the machine after an alarm
Press and hold for two seconds to spread the alarm window by 300 mm
Hg for arterial and venous pressures and fully open the transmembrane
(TMP) pressure window for 30 seconds. The light above the RESET key
will not be on.
During a blood leak alarm, press and hold for three seconds to override
the alarm and keep the blood pump running for three minutes. The light
above the RESET key will be on during an override.
Warning! During an override, the machine's blood leak
detector is inactive. You must monitor the treatment.
Note: The RESET key light flashes when a blood alarm
occurs.
2008T Machine Operator's Manual P/N 490122 Rev N
Press to
reactivate blood
system or reset
pressure alarm
limits as
prompted. Press
and hold to
temporarily widen
alarm limits or
override a blood
leak alarm.