2008T Machine Operator's Manual P/N 490122 Rev N
Function
The user can choose to do the Alarm Test with this button.
If the machine is set up with an automatic test valve, the user can choose
to do the Diasafe Test with this button.
The Verify Crit-Line button is used to manually verify the optional CLiC
device. To verify the CLiC device manually, clip the device on its
verification filter, select the Verify Crit-Line button, and press the
CONFIRM key. When an 'X' appears in the 'OK' field, the CLiC device
has been successfully verified.
Note: The 2008T hemodialysis machine is set up in Service
Mode to display either the "BTM BVM" screen or the "Crit-
Line" screen. The Verify Crit-Line button is available from
the "Test & Options" when the "Crit-Line" screen is displayed.
See the "Crit-Line Screen" on page 119 for more information.
Selecting the Patient ID button allows the user to enter a patient's ID in
the text box located to the right of the button. The 2008T hemodialysis
machine can upload treatment information to a network database for
review by clinical staff using a personal computer.
The High Flux button selects parameters for the use of a high flux
dialyzer for treatment. The selection is indicated by an X in the On or Off
box.
Warning! It is essential that the 2008T hemodialysis machine
balancing system is operating properly when using high-flux
dialyzers. The machine must successfully complete a
Pressure test before treatment commences. For more
information, See "Testing the 2008T Hemodialysis Machine"
on page 55.
The Arterial Width button allows the selection of three different ranges
for the arterial pressure alarm (120, 160, and 200 mm Hg).
Note: These options will only be available if set to "User Selectable" in
the Service Mode "Options" screen.
The Venous Width button allows the selection of four different ranges for
the venous pressure alarm (100 asymmetric limits, 120, 160, and 200 mm
Hg). The asymmetric limit will close the lower venous limit after a time
delay for stabilization.
Note: These options will only be available if set to "User Selectable" in the
Service Mode "Options" screen.
Runs the UF pump at preselected UF goal and time while recirculating.
UF goal and Time are entered in the Service Mode.
The Single Needle button prepares the machine for single-needle
dialysis treatment. For more information on single-needle dialysis
treatment, see Appendix A.
Chapter 3—Setting Treatment Parameters
93