Crit-Line Needs Verification; Crit-Line: No Blood - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message
Crit-Line needs
The CLiC device has not been
verification
verified within the past 30 days. The
Hct/BVCrit-Line graph on the "Crit-
Line" screen will be disabled.

Crit-Line: No Blood

The Tx Clock is running and the CLiC
device no longer senses blood in the
Blood Chamber.
185
Troubleshooting
Type
Action Required
Advisory
Verify the CLiC device:
message
If the message is not cleared:
If the message is still not cleared, the CLiC device cannot be used for
the dialysis treatment.
Warning
Check the Crit-Line Blood Chamber:
If the message is not cleared, the CLiC device cannot be used for the
dialysis treatment.
ALARMS
TROUBLE-
SHOOTING
Place the CLiC device on the CLiC Verification Filter, which is
attached to its USB cable.
Wait up to one minute for the message to clear. If the
message clears, the CLiC device is verified.
Select the "Test & Options" screen-button then select the
Verify Crit-Line button to start the verification process
manually.
Wait up to one minute for the message to clear.
Disconnect the CLiC device cable from the USB port.
Contact a qualified service technician.
Make sure there is proper blood flow: There should be no air
or bubbles in the chamber.
Make sure the CLiC device is properly seated on the Crit-
Line Blood Chamber: The device should be placed
perpendicularly over the chamber and cannot be rotated up
or down.
Remove the device from the Crit-Line Blood Chamber.
Disconnect the CLiC device cable from the USB port and
press the RESET key to disable CLiC device.
Contact a qualified service technician.
2008T Machine Operator's Manual P/N 490122 Rev N