Emptying Too Long; Error Reading Flash; Fail Actuator Temperature High/Low; Fail Actuator Venous High/Low - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

190
Troubleshooting

Emptying too long

The dialyzer empty program has
exceeded its maximum limit.

Error Reading Flash

Electronic Self Test
Fail * 9 Volt Battery
9V Power Failure Battery test has
failed.
Fail * Actuator Arterial
Arterial Pressure test has failed.
High
Fail * Actuator Arterial
Low
Fail * Actuator
Conductivity test has failed.
Conductivity High
Fail * Actuator
Conductivity Low
Fail * Actuator
Temperature test has failed.
Temperature High
Fail * Actuator
Temperature Low
Fail * Actuator TMP High
Transmembrane Pressure (TMP) test
has failed.
Fail * Actuator TMP Low
Fail * Actuator Venous
Venous Pressure test has failed.
High
Fail * Actuator Venous
Low
Fail * Air Detector
Air Detector test has failed.
2008T Machine Operator's Manual P/N 490122 Rev N
Alarm
If blood is not sensed, return the dialyzer supply and return lines to
the shunt and close the shunt door to terminate the program. If the
machine was in dialysis (blood sensed), turn machine power off and
back on to clear the program.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Test
Replace Battery
Message
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate if
failure repeats
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Reposition venous drip chamber. Rerun Test. If failure repeats,
Message
remove machine from service and alert a qualified service technician.