Fresenius Medical Care 2008T Operator's Manual page 31

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Feature
RS 232 Port
Volume
Control
(Dialysis
Mode Only)
Dialysis Mode
Speaker
Hour Meter
Ethernet Port
9-V Battery
(Power Failure
Alarm )
Heater On/Off
Switch
Mains Power
Supply On/Off
Switch
Note: Periodically check the power cord for damage (fraying, over-heating, cuts, scrapes,
etc.)
2008T Machine Operator's Manual P/N 490122 Rev N
Function
Electrically isolated RS 232 serial interface connector; hard wired, used
for display terminal connection.
Used to adjust the volume (sound pressure level) of the dialysis machine
audible warnings and alarms. Warning sound is adjustable between 75
dBA and 89 dBA (at 1 meter). Alarm sound is adjustable between 67 dBA
and 81 dBA (at 1 meter). Does not affect the volume from the separate
CDX speaker.
The Dialysis Mode speaker makes two different sounds: a warning
sound, and an alarm sound. The two sounds are distinct; the first one is
used for lower priority alarms, and the second for more important alarms.
This displays the number of hours the machine has run over its lifetime.
See the "2008T Preventive Maintenance Procedures Manual" (P/N
508033) for information on scheduled maintenance.
10/100 Ethernet connection for the CDX PC (optional); electrically
isolated.
9-V heavy duty alkaline battery for main power failure. A steady, audible
alarm will immediately sound for seven minutes that cannot be silenced
with the Mute key. It can be silenced manually, however, by removing
this 9-volt battery. See Replacing the 9-Volt Battery" on page 218 for
more information.
This switch turns the power to the dialysate heater on or off. This switch
must be in the ON position during treatment.
This switch turns the power to the whole machine on or off. This switch
must be in the ON position (|) to operate the machine.
Chapter 1—Overview
31