Troubleshooting

Resetting, Try Again.

Blood pressure module resetting
Retry > Press = XXX
The cuff pressure is too low to
measure the blood pressure. The cuff
pressure is XXX mmHg.

Reverse bloodlines

This message is a prompt for the
operator reverse the bloodlines for the
Access Flow test
Rinse Cond High
The Reverse Osmosis (RO) water
inlet conductivity is too high.
RO Water Cond High
The Reverse Osmosis (RO) water
inlet conductivity is too high.

RTD = Zero

The RTD (Remaining Time on
Dialysis) clock has counted down to
zero.
Running Diasafe Test
This message is advising the operator
of the status of the Diasafe test
Running Online PHT
This message is displayed when the
online Pressure Holding Test is in
progress.

Run Access Flow?

This message is a prompt for the
operator begin the Access Flow test.
Select Concentrate
This message is a prompt for the
operator to select a concentrate.
Select new goal or rate
UF time is out of range.
2008T Machine Operator's Manual P/N 490122 Rev N
Warning
Wait until blood pressure module completes resetting and retry blood
pressure reading.
Dialog
No action necessary
Message
Warning
To proceed with the Access Flow test, reverse the bloodlines and
press CONFIRM.
Opening
Press the RESET key to clear the message. Perform a Rinse cycle. If
Screen
alarm is not cleared, take the machine out of service and alert a
Message
qualified service technician.
Opening
Press the RESET key to clear the message. Perform a Rinse cycle. If
Screen
alarm is not cleared, take the machine out of service and alert a
Message
qualified service technician.
Warning
Reset the alarm. This message has alerted the operator that the
preset time on dialysis has elapsed (RTD = 0:00). If prescribed
treatment time has not been completed, the operator must take
further action to comply with the prescribed treatment.
Advisory
Advisory only. No action is required.
Advisory
Advisory only. No action is required.
Warning
To proceed now, press CONFIRM. To delay, press Escape. To
cancel, go to the "Kt/V AF" screen and toggle the Access Flow check
box to Off.
Advisory
To select a concentrate from the menu, use the  or  (Up/Down)
keys on the Keyboard to highlight the desired concentrate, and press
CONFIRM. For more information, see Chapter 3, "Setting an
Acid/Bicarbonate Type."
Dialog
Enter a new UF Goal or reduce UF time.